    Process gas analyzer - KM2000 CnHm

    The modular constructed ADOS KM 2000 CnHm EM equipment incorporates a microcontroller-aided measurement device for measuring solvents. All combustible gaseous CnHm compounds can be measured with the exception of chlorinated and sulphur-sublimed hydrocarbons.
    • 技術(shù)參數(shù)
    • 產(chǎn)品特點
    • 應用領(lǐng)域
    • 資料下載
    Measurement principle: Measuring the heat of combustion in a catalytic converter
    Measuring ranges: 0 - 50 mg C/m3 TOC up to 0 - 1600 g/m3 TOC
    Minimum detection limit: 1 mg C/m3 TOC
    Cross sensitivity
    (50 mg/m3 measuring range):
    concentration: max. deviation:
    200 mg/m3 SO2 -10 %
     
    30 mg/m3 NO2 -2,5 %
     
    300 mg/m3 CO +108 %
    300 mg/m3 NO +7 %
    Output signals: Current interface 0-(4)-20 mA
    max. load 400 ohm; RS 232
    Response time (t90): < 200 sec. sampling pipe (approx. 11 m; dead time 10 sec.)
    Accuracy: < 2 % full-scale error
    Permissible ambient temperature: +5 °C to +40 °C
    Temperature dependency: < 5 % full-scale error
    Sampled gas flow: 125 l / h (±10 l/h)
    Preheating time: approx. 120 min.
    Maintenance interval: 4 weeks with auto-calibration
    1 week without auto-calibration
    Mains supply: 230 V / 50 Hz; 150 V / 60 Hz; 600 VA
    Dimensions (W x H x D): 600 x 478 x 500 mm
    Weight: approx. 43 kg
    Test certificate: TÜV approval according to the clean-air regulations ("TA-Luft");
    TÜV-report: 936 / 21 200 245

    The thermocouples used for measurements, in conjunction with applying the principle of heat reaction, offer the following advantages:
     

    • High degree of sensitivity
    • Good accuracy
    • Negligible drift of zero point
    • Over-range signals have no effect

    Gas measurement system

    The sampled gas is drawn in by a pump through a feed pipe (heated if required), to the reaction chamber, via a compensating filter, flow regulator and flow-through meter. The gas is warmed to a constant temperature by means of the heater coil and jacket and finally burned in a solid-matter catalytic converter. The difference in temperature before and after combustion, is used as the measurement signal that is prepared and evaluated by the microcontroller-aided analyser.


    The analyser functions on the principle of heat reaction.The difference in temperatures at the reference measuring point and the measuring point, is a directly-dependent variable of the component part of combustible substances in the gas. The reference measuring point is subjected to the heated non-burned gas mixture, whilst the second probe of the thermocouple pile measures the temperature of the burned gas. A load-independent current of 0-(4)-20 mA is available for connection to electric test meters, plotters and limit value monitors. An RS 232 or RS 485 interface is incorporated for data communication.
    The inclusion of a measured value integration provides the facility of forming the average value of measured quantities, continuously or over a prescribed period of time.




    The hydrocarbon measuring system ADOS KM 2000 CnHm EM consists of the following 19" rack units:
    • Reaction chamber with sensor and electronics
    • Gas suction system with or without constant heating for the feed pipes, with sampled gas pump, flow-through meter, flow regulator, flow monitor and filter.
    • Microcontroller-aided evaluation unit in 19"-system with application specific standard plug-in Euro-cards
    • Housing

     

    Supervision of industrial processes

    • KM 2000 CnHm EM:
      Measuring the emission of hydrocarbons, according to the German clean-air regulations
    • KM 2000 CnHm:
      Measuring solvent saturation
      Measuring the concentration of solvents



    Concerning room air (ventilation) monitoring
    A warning is issued at very low concentrations of poisonous gas thus preventing any danger to health.

    版權(quán)所有 ? 2004- 深圳市昂為電子有限公司 ONUEE / ONWARDS(HK) Ltd. All rights reserved. 粵ICP備09176163號/
    <nobr id="ohfgb"></nobr>
    <output id="ohfgb"><thead id="ohfgb"><dl id="ohfgb"></dl></thead></output>

      
      
      <sup id="ohfgb"><thead id="ohfgb"><input id="ohfgb"></input></thead></sup>
      <output id="ohfgb"></output>
    • <mark id="ohfgb"><thead id="ohfgb"><pre id="ohfgb"></pre></thead></mark>

      婷婷五月丁香狠狠撸 | 在线A片网站 | 丁香五月亚洲无码 | 欧美性爱日韩精品 | 999在线免费视频 | se综合| 91久久久久乱码 | 欧美黄色毛片 | 成人网站在线浏览 | 亚洲欧美国产毛片在线 |